SanjVIEW 7 for Infrared Camera (IR)
SanjVIEW 7 for Infrared Camera (IR)
(Source Section 11)
Opening the software and setting up a session:
- Open SanjVIEW v7.0 from the desktop; center the sample in Project Manager and focus on the target device.
- System Status — ensure all indicator lights are green.
- Select Camera to Use — Infrared (IR), Visible – Thermoreflectance (TR), Near Infrared (NIR). Press F1 to switch.
- Click "Temp Control" tab; toggle ON/OFF and view readings for Stage, Thermal Coupler (TC), and ROI. Set the setpoint (example: 75), then click "Stage State" to turn on the stage.
- Click the "Auto Focus" tab (Piezo system control); select objective magnification (example: 5X).
- Click "Enable Motion" ON and "Center Stage" ON to center the Piezo controller. (Microscope may need refocusing after centering.)
- Click the file browser icon; choose the save folder and click "Current Folder."
- Select the desired imaging operation.
Note: Transient Thermoreflectance is not available with the Infrared camera.
Steady State
Select "Steady State" from Project Manager.
11.1.1 Hardware
- Enter the desired Base Temp (°C) (example: 75) and click "SET."
- Turn Stage State ON.
- Set "Vo ON Voltage" to 7V.
11.1.2 Timing / Illumination
- Set Device On Time to 10.
- Set Image Averages (#cycle) to 10.
11.1.3 3D Alignment — Focus ROI: same procedure as 8.1.3.
11.1.4 3D Alignment — Capture Reference: same procedure as 8.1.4.
11.1.5 Start Measurement
- Click "START Measurement."
- Progress in System Status bar; "Turn OFF" after each cycle; "Loop Iterations" tracked (9–11 sufficient for TR).
- Under Save Path, confirm directory/name.
- Click "Save Every Cycle," then "Save Data" when ready. (Save before stopping.)
- Click "STOP MEASUREMENT," then "Return to Project Manager" or File → Quit.
Calibration
Select "Calibration" from Project Manager.
11.2.1 Hardware
- Set "Magnification" to match the objective (example: 5x).
- Set "High Temp" to 120°C and "Base Temp" to 75°C.
- Turn "Stage State" ON.
- Click "Set Low Temp" to initialize chuck temperature.
- Click "Use Chuck Sensor" for the embedded sensor, or leave off for an external thermal probe.
11.2.2 Timing/Illumination
- Set Stage Heating Time to 40.
- Set Image Averages to 50.
11.2.3 3D Alignment — Focus ROI: same procedure as 8.1.3.
11.2.4 3D Alignment — Capture Reference: same procedure as 8.1.4, with the Load Reference alternative.
11.2.5 Start Measurement
- Click "START Measurement."
- Progress shown in System Status; "Calibration Done" after each cycle; "Loop Iterations" tracked.
- Select a region on Thermal Image View to view the live Calibration Factor ("Mean CTR").
- Under Save Path, confirm directory/name (example:
C:\Data\Qualification\Training). - Click "Save Every Cycle," then "Save Data." (Save before stopping.)
- Click "STOP MEASUREMENT," then "Return to Project Manager" or File → Quit.
Failure Analysis (FA) Mode
11.3.1 Frame Rate Selection
- In Project Manager, set "Cam Selection (F1)" to INFRARED.
- Uncheck "INT Vo 20mA Range."
- File → Options.
- Software tab → set "FA Mode Pulse Width" to Long (15 Frames).
- Return to Project Manager and click "FA Mode."
11.3.2 Process
Confirms the device is biased and conducting current before using Movie Mode.
- Go to "Biasing" tab.
- Change Vo(V) to 7.000.
- Click "Update Values."
- Confirm "Peak Voltage (V)" and "Peak Current (I)" read near 7V and 70 mA for device 2 (Center Diode).
- Note: If current is lower than expected, confirm "INT VO 20mA Range" is unchecked.
11.3.3 Save Data
- File → Save Data.
- Choose the save folder via File Browser and click "Current Folder" (example:
C:\Data\Qualification\Training). - Enter the file name (example: "FA-file").
- Click "Save."
- Click "Return" to close the pop-up.
- Note: Change the Microsanj Data Save Path before saving new data, to avoid overwriting.
Movie Mode
Select "Movie Mode" from Project Manager.
11.4.1 Control Settings
- Set "Voltage to Device (V)" to 7V.
- Click "Update V."
- Set "DUT Pulse (s)" (example: 1s). Click "Update Timing."
- Confirm "Save Whole Series" is on (on by default).
- Click "File Path" icon (or type the path) and select a name.
- Click "Start Movie" on Movie Mode Acquisition.
- Click "Save Image."
- Note: Movie Mode saves a series of a 105-file cycle when "Save Image" is clicked.
SanjANALYZER
11.5.1 System Status
- Confirm Camera is set to INFRARED.
- Click SanjANALYZER to open the "SanjVIEW Image Analyzer" window.
11.5.2 Process Data — identical procedure to Section 8.5.2 (Calibration tab → Point to Point → Load CTR map → select analysis file → draw up to 5 ROI boxes → adjust scale/averaging → Settings tab to save the processed image → optional "Process Series" for a file series).