Calibration — Quick Start

Calibration — Quick Start

Calculates the thermoreflectance coefficient (TR) or emissivity (IR) of specific device surfaces. Accurate temperature readings depend on a coefficient obtained via calibration. Coefficient maps can be produced for point-by-point calibration. The module thermally cycles the DUT while reading the thermocouple back, continuously averaging coefficients across cycles to raise SNR.

Key features:

Verifying calibration using TCAT-100 in pulsed mode: set up a test sample as for calibration, launch Pulsed Mode software, select TCAT-100 as the output device, and enable thermocouple readback to simultaneously display the thermocouple signal alongside TR or IR measurement data. Calibrated IR camera results should closely overlap thermocouple results. Note: the thermocouple used with the TCAT module has a response time of ~0.5 s, so it may be necessary to lengthen the timescale to see an accurate response.