Project Manager

Project Manager

Launch SanjVIEW from the Windows 10 start screen shortcut; once the license verifies, the Project Manager window opens. From here you can select:

Key features of the Project Manager window:

Beginning thermal imaging on a new device:

  1. Place the device under test (DUT) on the probe station.
  2. In Project Manager, confirm all components show a green status light (unless the NIR camera is unavailable).
  3. Select the desired illumination option and "Channel Number," and focus the microscope. A live DUT image appears on the right of the screen.
  4. Check the "Auto" box under LED Brightness for automatic brightness optimization (Visible Camera only).

Note: When imaging with infrared light (IR), Transient Thermoreflectance options are unavailable, and LED illumination is disabled during an IR imaging session.

Viewing temperature response in DC mode: while Project Manager is running, approximate temperature values display on the camera measurement window, scrolling as an average of the drawn Region of Interest (ROI) (or the whole image if no ROI is drawn). Use the rectangle tool to draw an ROI. DUT bias controls let you select outputs and voltage while acquiring images; the TC-100 stage can be used (~100 mA/V), or the Relay output can be selected from the pulldown. Select the TC-100 tab and press "TC-100 State on" to read the thermocouple value (useful for setting IR camera gain). Right-click the image window to save the acquired image and its color palette. [diagram/image in source — not reproduced here: Figures 1–3, Project Manager and DC mode window screenshots]


Revision #1
Created 2026-08-26 18:16:48 UTC by Accounts