Transient Thermoreflectance (TR)

Transient Thermoreflectance (TR)

Acquires new thermal imaging data for pulses between 50 µs and 5 ms (transient).

Key features:

Example procedure — obtaining a transient measurement:

  1. Turn off "Int Vo 20mA Range" in the Project Manager.
  2. Set "DUT Pulse" to the desired pulse width (example: 500 µs).
  3. Set "Voltage to Device" (example: 6.0 V) and select "Update Values."
    1. Switch to "Thermal Imaging" mode.
    2. Confirm the device is heating as expected for the 500 µs delay.
  4. Set "Image Delay(s)" to 0 to initialize delay to 0 µs (you can start at any initial delay time).
  5. Go to File → Setup Sweep.
  6. Enter "Number Delay Images" (example: 51).
  7. Enter "Delay Steps(s)" (example: 10.0 µs) — software will take 51 images at 10 µs steps starting at 0 s (0–500 µs).
  8. Enter "Ave time per frame(s)" (example: 5 s per image).
  9. Enter the "Base File Name" to save under.
  10. Select an ROI with the "Draw ROI" tool.
  11. Press "Start."
  12. The software steps through the thermal transient automatically and shuts off voltage on completion. A thermal image is averaged and saved at the configured interval (example: every 5 s; 255 s total for 51 points, 4 min 15 s).
  13. Once done, exit and open the saved data in SanjANALYZER.

The transient response of an ROI can be watched live in the "Region Mean Graph" tab (lower-left) by selecting a region with "Draw ROI."


Revision #1
Created 2026-08-26 18:16:49 UTC by Accounts