Calibration
Calibration
Perform the common setup in Quick Start Steps, then click Calibration to start. Continue with the steps below.
- Hardware
- Set Base Temp to 20 °C and click Set Low Temp.
- Note: The temperature set on the 4D Nano Align should change. If it does not, make sure the Chiller is on.
- Set High Temp to 120 °C.
- Note: Microscope Settings are auto-filled from Project Manager (no need to reselect them).
- Set Base Temp to 20 °C and click Set Low Temp.
- Timing/Illumination
- Set Stage Heating Time to 50.
- Set Image Averages to 55.
- Auto-select the Brightness (%).
- Note: Brightness % is auto-filled from Project Manager (no need to reselect it).
- 3D Alignment – Focus ROI
- Select a region on the CCD image using the rectangle tool. Choose a unique feature on the sample, such as edges or scratch marks.
- Press Focus ROI to set the selected region as your ROI.
- Wait a few seconds until the selected region is focused as your ROI and the Focus ROI indicator turns green.
- Note: The selected ROI region displays in green. To zoom in/out, use the magnifying glass — hold Shift to zoom out, Shift+Click to zoom in.
- 3D Alignment – Capture Ref
- After selecting the ROI, keep the same ROI or select another region with more distinct features for alignment.
- Press Capture Ref to set the newly selected region as the Reference region.
- Wait for the Capture Ref indicator to turn green.
- Set the initial parameters for Focus Tolerance, Alignment %, and Vibration Filter. The indicator lights turn green when the image is in range.
- Note: Increasing the Alignment % value and decreasing the Vibration Filter improves alignment accuracy.
- Start Measurement
- Once setup is complete, click START Measurement to begin calibration.
- Save Images
- The calibration cycles through temperatures to reduce noise and improve accuracy. Typically after 3 cycles, the thermoreflectance error is <1E-5 (or <5% for 530 nm on gold material). The System Status bar shows progress and displays "Calibration Done" after each cycle, and counts cycles under "Loop Iterations".
- Note: Each cycle takes about 4 minutes.
- Select a region on the Thermal Image View to get the Thermoreflectance Coefficient (Cth) while the measurement is running. (Example in source: the middle of the blue 5 µm x 500 µm band is selected.)
- Under Save Images, check that the directory and "Save Path" name are correct.
- Check the Save Every Cycle box. When ready to save the data, click Save Image.
- Note: Save the image before stopping the measurement.
- To stop the measurement, click STOP MEASUREMENT, or click File > Quit.
- The calibration cycles through temperatures to reduce noise and improve accuracy. Typically after 3 cycles, the thermoreflectance error is <1E-5 (or <5% for 530 nm on gold material). The System Status bar shows progress and displays "Calibration Done" after each cycle, and counts cycles under "Loop Iterations".