Steady State Image
Steady State Image
Perform the common setup in Quick Start Steps, then click Steady State to start. Continue with the steps below.
-
Hardware
- Set the Base Temp to 20 °C.
- Set Device ON Voltage to 5 V and Device OFF Voltage to 0 V.
- Note: Microscope Settings are auto-filled from Project Manager (no need to reselect them).
-
Timing/Illumination
- Set Device On Time to 10.
- Set Image Averages to 50.
- Set the Duty Cycle to 50%.
- Auto-select the Brightness (%).
-
3D Alignment – Load Ref
- Click Load Ref.
- Select the saved file from Calibration.
- Wait a few seconds until the selected region is focused and the Load Ref indicator turns green.
- Set the initial parameters for Focus Tolerance, Alignment %, and Vibration Filter. The indicator lights turn green when the image is in range.
- Note: Increasing the Alignment % value and decreasing the Vibration Filter improves alignment accuracy.
-
Start Measurement
- Once setup is complete, click START Measurement to begin the measurement.
- Note: the source document literally reads "to begin calibration" here; this appears to be a copy/paste artifact carried over from the Calibration section rather than a distinct step — flagged here for anyone updating this procedure against the vendor source.
- Once setup is complete, click START Measurement to begin the measurement.
-
Save Images
- After sufficient averaging time, the hot line should appear (a few seconds).
- Select a region on Thermal Image View to get the ROI Temperature Value. (Example in source: the middle of the green 5 µm x 500 µm band is selected.)
- Enter the estimated Thermoreflectance Coefficient (Cth) from the calibration to get a more accurate ROI Temperature.
- Under Save Images, check that the directory and "Save Path" name are correct.
- When ready to save the data, press Save Image.
- Note: Save the image before stopping the measurement.
- To stop the measurement, click STOP MEASUREMENT, or click File > Quit.
Note on source step numbering: the vendor source labels both "Click START Measurement" and "Save Images" as step "4" in this section. They are presented here as steps 4 and 5, in the order they appear in the source, for clarity.