5. Verify Calibration Using TCAT-100 in Pulsed Mode 5. Verify Calibration Using TCAT-100 in Pulsed Mode To verify calibration coefficients obtained from the Calibration module: Set up a test sample the same way as for a standard calibration run. Launch the Pulsed Mode software and select the TCAT-100 as the output device. Select to read back the thermocouple. This lets the software simultaneously display the thermocouple signal alongside the Thermoreflectance or Infrared measurement data. Compare results: the calibrated IR camera results should closely overlap the thermocouple results. [diagram/image in source — not reproduced here: Figure 13, pulsed-mode operation on the Microsanj test sample; Figure 14, final calibrated measurement in Pulsed Imaging mode] Note: The thermocouple used with the TCAT module has a response time of approximately 0.5 seconds, so it may be necessary to lengthen the time scale to view an accurate response.