SanjANALYZER™ — Thermal Image Analyzer
SanjANALYZER™ — Thermal Image Analyzer
A fully integrated analysis package for low-frequency and transient thermal imaging data. Capabilities:
- View previously acquired thermal images
- Correct for thermoreflectance coefficients
- Utilize point-by-point calibration maps
- Find area mean temperature values
- Create and save cross-section plots
- Output images in various formats
- Create .avi movies from an image series (SanjANALYZER Plus only)
Key features:
- Load a Thermal Image — low-frequency data uses
.fdi; transient data uses.tti. Double-click a file in the file selection control to load the CCD and thermal images. - CCD Image Display — grab, zoom (SHIFT-CLICK to zoom out), area select tools.
- Primary Thermal Image Display — viewing modes: Thermal Image VIEW, Merge VIEW, and optional 3D VIEW (SanjANALYZER PLUS).
- Thermal Image VIEW — calculation using masking info or a single coefficient. ROI Mean/Statistics: select a shape or freehand region for mean temperature (selection also shows on the CCD image). Cross-Section Graph: draw a line to plot temperature vs. distance (masked/invalid points excluded); select the correct Objective for accurate scale; save in ASCII (tab-delimited). Limit Bars: coerce auto-scaled data to selected limits (also affects Merge and 3D views). Color Palette Selection: multiple palettes, saved with exported images. Software Binning (x2 or x4): spatial averaging to improve faint images or cross-section SNR, at the cost of spatial resolution.
- Merge VIEW — for display/saving only (temperature values live only in the thermal image). Converts to RGB for compatibility with image output formats. CCD/Thermal/Thermal Image slide bars: first slider overlays % of CCD image on the thermal image (to locate hotspots precisely); gain/offset sliders apply to the thermal image (100%/0% = unchanged).
- Crop from CCD Image checkbox — crop the thermal image away from the ROI, useful for reducing .avi movie file size.
- Fix Temperature Scale checkbox — keeps a previously set scale fixed across a loaded series (recommended workflow: find the hottest image in the series, set the desired scale, check "fix scale," return to the first image, and run "process series").
- Show Voltage / Delay / Color bar checkboxes — overlay acquisition conditions and the color-to-temperature mapping on the image.
Example procedure — analyzing a transient thermoreflectance image series:
- Select the first image of the transient series.
- Enter the "number of images in the series" (example: 51).
- Select an ROI box on the area of interest (example: a 5 µm gold heater).
- Select "Process Series." The software steps through the series and plots data in "ROI Mean Series" (bottom left); right-click the y-axis to autoscale.