4. Obtain a Movie Mode Image
4. Obtain a Movie Mode Image
The Movie Mode module acquires thermal imaging data for pulses longer than 5 ms (quasi steady state); it is IR-only. [diagram/image in source — not reproduced here: Figure 7, Movie Mode Window]
Optional configuration (before running):
- Go to Options > Software and use the "Movie Mode Length" drop-down to set the averaging time window, 3–30 seconds. [diagram/image in source — not reproduced here: Figure 8, time-window control]
Worked example:
- Connect the Device Under Test (DUT) to Voltage Out (the AUX Output is optional).
- Select the biasing conditions for the DUT via the "Device Control" panel and click "Update Values."
- While the image is averaging, use the ROI tool, the scale bars, and the Thermal Image Display tool to optimize the image.
- Once the desired thermal image is achieved through averaging, save a data file for further analysis and output.
Reference: Voltage Out (Output 1) is the primary, camera-synced output always used for thermal imaging; Output 2 is a constant secondary source that can power other parts of the circuit. Neither output changes until "Update Values" is clicked.