5. Verify Calibration Using TCAT-100 in Pulsed Mode
5. Verify Calibration Using TCAT-100 in Pulsed Mode
To verify calibration coefficients obtained from the Calibration module:
- Set up a test sample the same way as for a standard calibration run.
- Launch the Pulsed Mode software and select the TCAT-100 as the output device.
- Select to read back the thermocouple. This lets the software simultaneously display the thermocouple signal alongside the Thermoreflectance or Infrared measurement data.
- Compare results: the calibrated IR camera results should closely overlap the thermocouple results. [diagram/image in source — not reproduced here: Figure 13, pulsed-mode operation on the Microsanj test sample; Figure 14, final calibrated measurement in Pulsed Imaging mode]
Note: The thermocouple used with the TCAT module has a response time of approximately 0.5 seconds, so it may be necessary to lengthen the time scale to view an accurate response.