SanjVIEW 7 for Near Infrared (NIR) Camera

Nano Therm System Operation

SanjVIEW 7 for Near Infrared (NIR) Camera SanjVIEW 7 provides the same functionality for Visible and NIR cameras, provided the camera selection (F1: VISIBLE/NIR) is compatible with the LED selection (F3). See the Camera/LED Compatibility table in Specification...

SanjVIEW 7 for Infrared (IR) Camera

Nano Therm System Operation

SanjVIEW 7 for Infrared (IR) Camera Opening and initial setup: Open SanjVIEW v7.0 from the Computer desktop. Center the sample in Project Manager and focus on the desired device. System Status: Confirm all indicator lights are green for your connected system. ...

EZ530 Therm System

Overview

EZ530 Therm System

Overview The EZ530 Therm system (referred to throughout the source manual as the EZ-THERM Series / Eazy-THERM) is a benchtop, non-contact thermal imaging and analysis platform built around Thermoreflectance (TR) microscopy, with optional Infrared (IR) and Near...

Safety Precautions

EZ530 Therm System

Safety Precautions (Source Section 1.1–1.2) High temperature: Do not touch the thermal plate while it is in operation — contact may cause severe burns. No user-serviceable parts inside the EZ-THERM controller. Do not remove covers. Exposed AC power can be pres...

Specifications

EZ530 Therm System

Specifications NT220 Platform Specification Table (Source Section 5, "Nano THERM Specification." The EZ530 corresponds to the PKG-2 column — see Overview note above.) Parameter Base PKG-1 PKG-2 Transient Response 50 ns LED Pulse Duration (FWHM), 12.5 ns Delay ...

Setup and Installation

EZ530 Therm System

Setup and Installation Connecting the EZ-THERM System (Source Section 4. Ensure all modules are powered off before making any connections. Complete all connections and power on all modules before running the SanjVIEW 7 software.) 4.1 Power up Eazy-THERM. Befor...

Maintenance

EZ530 Therm System

Maintenance Chiller maintenance (Source Section 3.11): Replace chiller liquid every 3 months under light usage, or monthly under heavy usage. Use distilled water only; drain before extended non-use or transport. Keep the unit ≥30 cm from obstructions at the re...

Appendix — Thermoreflectance Technical Reference

EZ530 Therm System

Appendix — Thermoreflectance Technical Reference (Source Appendix A/B — retained here as background reference material.) Thermoreflectance Imaging Fundamentals Thermoreflectance imaging is based on a linear change in a material's reflection coefficient with te...

Source Documents

EZ530 Therm System

Source Documents This reference was compiled from: EZ530 Therm User Manual [source-pdf].txt — text extraction of the Microsanj LLC "Manual User Guide, EZ-THERM Series" (Copyright 2025), located at: /tmp/claude-1000/-home-davidspitzer-Android-Auto-replacement/f...

Operation

EZ530 Therm System

SanjVIEW 7 Software — Overview

EZ530 Therm System Operation

SanjVIEW 7 Software — Overview (Source Section 7) The EZ-THERM General Purpose Thermal Analyzer is a Thermoreflectance (TR) system: a benchtop precision analysis tool for static and transient thermal imaging of active semiconductors from millimeter to micron s...

Operating SanjVIEW 7 with the Visible Camera (TR)

EZ530 Therm System Operation

Operating SanjVIEW 7 with the Visible Camera (TR) (Source Section 8) Opening the software and setting up a session: Open SanjVIEW v7.0 from the computer desktop; center the sample in Project Manager and focus on the target device. System Status — ensure all in...

SanjVIEW 7 for Near Infrared (NIR) Camera

EZ530 Therm System Operation

SanjVIEW 7 for Near Infrared (NIR) Camera (Source Section 10) A visible camera (mono or color) captures images in the visible light spectrum (400–780 nm), producing images resembling human vision. A NIR camera captures near-infrared light (780–1400 nm), invisi...

SanjVIEW 7 for Infrared Camera (IR)

EZ530 Therm System Operation

SanjVIEW 7 for Infrared Camera (IR) (Source Section 11) Opening the software and setting up a session: Open SanjVIEW v7.0 from the desktop; center the sample in Project Manager and focus on the target device. System Status — ensure all indicator lights are gre...

Device Biasing, Trigger Function, and Timing

EZ530 Therm System Operation

Device Biasing, Trigger Function, and Timing (Source Section 9) 9.1 Diagram for TR Imaging — SanjCONTROLLER PLUS™ signals: A. VI EXT: Input for fixed DC voltage up to +60V. B. VO EXT: Pulsed output of the DC voltage applied to VI EXT. C. AUX INT: Up to ±10V DC...

NT220 System Quick Start

Overview

NT220 System Quick Start

Overview The NT220 is Microsanj's transient thermoreflectance imaging system (Visible & NIR Camera configuration), used to generate thermal images and temperature measurements of microelectronic devices under test. The system combines: A visible or NIR camera ...