Notes and Cautions
Notes and Cautions Notes and cautions called out throughout the source procedures, consolidated here for reference: Stage clearance: Before using Motion Control (Center Stage / Enable 3D Motion), make sure the stage is clear. Chiller check: During Calibration ...
Source Documents
Source Documents Quick Start Guide for NT220 System – Visible & NIR Camera – Microsanj Test Sample, Microsanj, LLC, Copyright 2021 (source PDF, text-extracted). Microsanj, LLC, 48421 Milmont Drive, Fremont, CA 94538; info@microsanj.com.
Overview
Overview This document is a cleaned-up, internal reference version of the "SanjVIEW Quick Start" guide (Microsanj LLC, 48421 Milmont Drive, Fremont, CA 94538; authored by Huy Q Le; source revision Rev241029). Despite the "Quick Start" title, the source is a fu...
Source Documents
Source Documents SanjVIEW Quick Start (source PDF, text-extracted) — SanjVIEW Quick Start TEA Sample [source-pdf].txt Microsanj LLC, 48421 Milmont Drive, Fremont, CA 94538, (408) 470-7539. Author: Huy Q Le. Document revision: Rev241029. Covers: SanjVIEW Softw...
1. Launch SanjVIEW and Prepare for Imaging (Project Manager)
1. Launch SanjVIEW and Prepare for Imaging (Project Manager) Launch SanjVIEW from the Windows Start screen shortcut. Wait for the software license to be verified — the Project Manager window opens. [diagram/image in source — not reproduced here: Figure 1, Proj...
2. View Temperature Response in DC Mode
2. View Temperature Response in DC Mode While Project Manager is running, use the rectangle tool to draw a Region of Interest (ROI) on the sample in the image display window. (If no ROI is drawn, the entire image is averaged.) The approximate temperature for t...
3. Acquire a Transient Thermoreflectance (TR) Measurement
3. Acquire a Transient Thermoreflectance (TR) Measurement The Transient Thermoreflectance module acquires thermal imaging data for pulses between 50 µs and 5 ms; it is only available with the Visible Camera. Worked example (values as given in the source): Turn...
4. Obtain a Movie Mode Image
4. Obtain a Movie Mode Image The Movie Mode module acquires thermal imaging data for pulses longer than 5 ms (quasi steady state); it is IR-only. [diagram/image in source — not reproduced here: Figure 7, Movie Mode Window] Optional configuration (before runnin...
5. Verify Calibration Using TCAT-100 in Pulsed Mode
5. Verify Calibration Using TCAT-100 in Pulsed Mode To verify calibration coefficients obtained from the Calibration module: Set up a test sample the same way as for a standard calibration run. Launch the Pulsed Mode software and select the TCAT-100 as the out...
6. Analyze a Transient Thermoreflectance Image in SanjANALYZER
6. Analyze a Transient Thermoreflectance Image in SanjANALYZER SanjANALYZER opens .fdi (Low Frequency) or .tti (Transient) data files. Load a thermal image by double-clicking the file of interest in the file selection control; the CCD and thermal images then...
7. SanjANALYZER Plus — Calibration Image and ΔT Measurement
7. SanjANALYZER Plus — Calibration Image and ΔT Measurement 7.1 Open Calibration Image Open the file where all test images are saved. Open the Calibration Image file. Click "Correct XY" to correct for thermal expansion. 7.2 Save Calibration Image Click "Save I...
8. Visible Camera Operating Modes (Reference Screens)
8. Visible Camera Operating Modes (Reference Screens) Source section 9 walks through the same acquisition modules described above, screen-by-screen, using the Visible Camera. The source provides screenshots for each mode with little or no additional body text;...
9. Infrared (IR) Camera Workflow
9. Infrared (IR) Camera Workflow 9.1 System Status and Initial Setup Open SanjVIEW v7.0 from the desktop, then center the sample in the Project Manager to focus on the desired device. Ensure all indicator lights are green for the connected system. Select the c...
10. Device Biasing, Trigger Function, and Timing Reference
10. Device Biasing, Trigger Function, and Timing Reference SanjCONTROLLER PLUS™ I/O (for TR imaging): Label Description A. VI EXT Input for fixed DC voltage up to +60V B. VO EXT Pulsed output of DC voltage input to VI EXT C. AUX INT Up to ±10V DC for secondary...